Analytical non-conformities in bacteriology : Incidence and impact

##plugins.themes.academic_pro.article.main##

Siwar Frigui
Salim Ben Jaafer
Yosra Chebbi
Anis Raddaoui
Wafa Achour

Abstract

Introduction: The incidence of analytical non-conformities (NC) is a particularly interesting quality indicator (QI) for evaluating the analytical process in conventional bacteriology. Our aims were to determine the incidence of analytical NC in bacteriology and to assess their impact on the time to results and on the cost.
Methods: We conducted a prospective descriptive study including all samples received at the Laboratories Department of the National Bone Marrow Transplant Centre for bacteriological analysis, over a 5-month period (July-November 2023).
Results: In total, 180 analytical NC were recorded, with an incidence of 56.7/1000 samples and 30.4/1000 technical operations. The number of NC was correlated with both the number of samples (R=0.564; p=0.006) and the number of technical operations (R=0.471; p=0.027). Analytical NC were dominated by incorrect culture (29%), error in antibiotic susceptibility testing (25%) and non-completion of a technical operation (20%). Causes of NC were dominated by the lack of training (28%) and the distraction (22%). Out of 180 analytical NCs reported, 102 had an impact on the
time taken to deliver results (56.7%), estimated at one day in all cases, and 77 had an impact on the cost of reagents (42.8%), estimated at 1056TD during the study period.
Conclusion: This study showed a high incidence of analytical NC in bacteriology, closely linked to the number of samples and technical operations, with a significant impact on the time to results and the cost of analyses.

Keywords:

Non-conformity , Quality , Accreditation , Bacteriology , Management

##plugins.themes.academic_pro.article.details##

References

  1. Guzel O, Guner EI. ISO 15189 Accreditation: Requirements for quality and competence of medical laboratories, experience of a
  2. laboratory I. Clin Biochem. 1 mars 2009;42(4):274-8.
  3. Allen LC. Role of a quality management system in improving patient safety — Laboratory aspects. Clin Biochem. 1 sept
  4. ;46(13):1187-93.
  5. Laudat P, Galinier JL, Cattoen C, Ferroni A, Lamy B, Courcol R. L’accréditation en bactériologie : enjeux, difficultés et particularités.
  6. Rev Francoph Lab. 1 avr 2014;2014(461):25-30.
  7. International Organization for Standardization. ISO 15189:2022 Medical laboratories -Requirements for quality and competence.
  8. th ed. Geneva: International Organization for Standardization; 2022
  9. Dauwalder O, Landrieve L, Laurent F, de Montclos M, Vandenesch F, Lina G. Does bacteriology laboratory automation reduce time to
  10. results and increase quality management? Clin Microbiol Infect Off Publ Eur Soc Clin Microbiol Infect Dis. mars 2016;22(3):236-43.
  11. Giannoli JM, Szymanowicz A. Propositions de recommandations pour l’utilisation pratique des contrôles internes de qualité (CIQ)
  12. dans un laboratoire de biologie médicale. Ann Biol Clin (Paris). 1 août 2011;69(4):489-98.
  13. Jnah A, Yagoubi M, Seffar M, El Hamzaoui S, Hamamouchi J, Zouhdi M. Control of non-conformities in the pre-analytical phase at the
  14. Bacteriology Laboratory of the Ibn Sina University Hospital in Rabat (Morocco). Tunis Med. 1 mars 2022;100(3):247-54.
  15. Hani YE, Alaoui HL, Mahassini N, Radi L, Zouhdi M, Filali-Maltouf A. Risk analysis and management of non-conformities of the pre-
  16. analytical phase in a university testing laboratory of bacteriology. J Med Lab Diagn. 31 janv 2014;5(1):1-10.
  17. Société Française de Microbiologie. REMIC : Référentiel en microbiologie médicale. 7ème édition. Paris : Société Française de
  18. Microbiologie; 2020
  19. Comité de l'Antibiogramme de La Société Française de Microbiologie. Recommandations 2024 V.1.0 Juin. Paris : Société
  20. Française de Microbiologie; 2024
  21. Ngubo MA. The effect of human factors on implementation of accreditation in the medical laboratories of KwaZulu-Natal. [South
  22. Africa]: Durban University of Technology;
  23. Bouzid K, Zrelli S, Rebeh A, Bartkiz A, Khlifi A, Jerbi YE, et al. Preanalytical phase management: identification and control of
  24. nonconformities in the sampling room of a clinical laboratory in Tunisia. J Lab Precis Med [Internet]. 18 avr 2018 [cité 10 déc
  25. ;3(4). Disponible sur: https://jlpm.amegroups.org/article/ view/4330
  26. Kritikos A, Prod’hom G, Jacot D, Croxatto A, Greub G. The Impact of Laboratory Automation on the Time to Urine Microbiological
  27. Results: A Five-Year Retrospective Study. Diagnostics. janv 2024;14(13):1392.
  28. Plebani M. The detection and prevention of errors in laboratory medicine. Ann Clin Biochem. 1 mars 2010;47(2):101-10.
  29. Barbé B, Yansouni CP, Affolabi D, Jacobs J. Implementation of quality management for clinical bacteriology in low-resource settings. Clin Microbiol Infect. 1 juill 2017;23(7):426-33.
  30. Miédougé M, Izopet J. Une expérience pratique d’accréditation en virologie. Rev Francoph Lab. 1 févr 2010;2010(419):65-8.
  31. Nassiri O, Bousnina F, Hami A. Non-conformities in the Pre-analytical Phase at the Parasitology-Mycology Laboratory of the Mohammed VI University Hospital in Oujda. Cureus [Internet]. 18 juin 2024 [cité 11 févr 2026]; Disponible sur: https://www.cureus.com/
  32. articles/246812-non-conformities-in-the-pre-analytical-phase- at-the-parasitology-mycology-laboratory-of-the-mohammed-vi-
  33. university-hospital-in-oujda
  34. Qin J, Zhang H, Zhang X, Wang L, Yu Y, Li M, et al. Optimizing blood culture diagnostics through laboratory automation: reducing
  35. turnaround time and improving clinical outcomes. Microbiol Spectr. 11 nov 2025;13(12):e01927-25.
  36. Bergaoui, H., Belhedi, M., Lazzem, W., Bacha, O., Ba, A., Ayoub, M., & Chouaieb, S. (2025). Pre-analytical non-conformities in
  37. biochemistry: frequency and economic burden. Poster presented at the 7th EFLM Conference on Preanalytical Phase, Padova, Italy